Karl Fischer & Potentiometric Titration Application Library

Copper Analysis in Copper Sulfate Plating Baths Using Redox Titration
Copper concentration is critical to plating performance and coating quality. Learn how automated redox titration improves accuracy, repeatability, and process control in copper sulfate baths.

Nickel (Ni²⁺) Analysis in Nickel Plating Baths Using EDTA Titration
Nickel concentration is the most critical parameter in plating bath control. Learn how automated EDTA titration improves accuracy, repeatability, and coating consistency.

Determination of hydrogen peroxide and sulfuric acid in etching solution | Autotitrator COM-A19
1) Determination of sulfuric acid by neutralization titration with sodium hydroxide・・・(1)
2) Determination of hydrogen peroxide by redox titration with potassium permanganate standard solution.・・・ (2)
H₂SO₄ + 2NaOH → Na₂SO₄ + H₂O・・・(1)
5H₂O₂ + 2KMnO₄ + 3H₂SO₄ → K₂SO₄ + 2MnSO₄ + 8H₂O + 5O₂・・・(2)

Determination of Tin (Sn²⁺) in solder plating solution | Autotitrator COM-A19
Control active tin (Sn²⁺) levels in solder plating baths to improve coating quality and prevent oxidation-related process instability.

Chromium (Cr⁶⁺ and Cr³⁺) Analysis in Plating Baths for Process and Environmental Control
Control hexavalent and trivalent chromium levels in plating baths to improve coating performance, reduce defects, and support environmental compliance.

Nickel Chloride Analysis in Nickel Plating Baths for Process Control
Nickel chloride plays a critical role in plating bath conductivity and anode performance. Learn how accurate analysis improves process stability and coating quality.

Determination of alkalinity in bottled water | Autotitrator COM-A19
Total amount of the hydroxides and half amount of carbonates are measured when it is titrated to about pH 8.3 with sulfuric acid titrant.
2OH⁻ + H₂SO₄ → SO₄²⁻ + 2H₂O
2CO₃²⁻ + H₂SO₄ → SO₄²⁻ + 2HCO₃⁻
All of the bicarbonates are neutralized when it is titrated successively to about pH 4.8.
HCO₃⁻ + H₂SO₄ → SO₄²⁻ + 2CO₂ + H₂O
This report introduces a measurement example for bottled water using potentiometric titration as end point detection method according to Standard Methods for the Examination of Water.

Determination of residual chlorine in tap water | Autotitrator COM-A19
Standard Methods for the Examination of Water adopts the following methods as determination method for residual chlorine.
・Dimethyl -p- phenylenediamine (DPD) method ………………………………………………Colorimetric method
・Dimethyl -p- phenylenediamine (DPD) titration method………………………………………Titration method
・Amperometric titration method (Titration method with phenylarsenoxide solution) ………Titration method
・Iodometric titration method………………………………………………………………………Titration method
This report introduces an example for determination of residual chlorine in tap water with iodometric titration method.
Residual chlorine oxidizes potassium iodide in acidic solution to generate free iodine.
Cl₂ + 2I⁻ → I₂ + 2Cl⁻
The generate iodine is potentiometrically titrated with sodium thiosulfate standard solution.
I₂ + 2Na₂S₂O₃ → 2NaI + Na₂S₄O₆

Calcium hardness in tap water | Autotitrator COM-A19
The measurement method for calcium hardness is similar to that of total hardness. Adjust the pH of sample water to higher than pH 12 with sodium hydroxide to mask the reaction of magnesium and EDTA. Start titration with 2-hydroxy-1-(2-hydroxy-4-sulfo-1-naphthylazo)-3-naphthoic acid (NN) indicator. Its color changes from red to blue.
CaCO₃ + Na₂EDTA → CaEDTA + Na₂CO₃
This report introduces an example of calcium hardness measurement for tap water with photometric titration method using EDTA standard solution according to the Standard Methods for the Examination of Water.

Total hardness in tap water | Autotitrator COM-A19
(1) Total hardness Total amount of calcium and magnesium ions
(2) Calcium hardness Calcium ion
(3) Magnesium hardness (Total hardness) – (Calcium hardness)
(4) Non-carbonate hardness (Permanent hardness)
(5) Carbonate hardness (Temporary hardness)
These are stipulated in some applicable standard such as Standard Methods for the Examination of Water, Standard Methods of Analysis for Hygienic Chemists, and JIS K0101 Testing methods for industrial water etc.
This report introduces an example for determination of total hardness in tap water with photometric titration method using EDTA standard solution according to the Standard Methods for the Examination of Water.
Take 100 mL of sample and add 1 mL of 0.01 mol/L magnesium chloride, 2 mL ammonia buffer, and 0.2 mL of EBT indicator. Titrate with 0.01 mol/L EDTA standard solution (red →blue color). Perform the same procedure for 100 mL of DI water instead of sample as blank measurement.
CaCO₃ + Na₂EDTA → Ca-EDTA + Na₂CO₃
MgCO₃ + Na₂EDTA → Mg-EDTA + Na₂CO₃

Quantitative determination of chloride ions in tap water | Autotitrator COM-A19
100 mL of the sample water is collected and acidified with nitric acid for potentiometric titration with silver nitrate titrant.
Cl⁻ + AgNO₃ → AgCl↓ + NO₃⁻

Standardization of benzethonium chloride standard solution | Autotitrator COM-A19
This report introduces a method for standardization of benzethonium chloride standard solution. Sodium dodecyl sulfate is used for standard solution. First, determine the purity of sodium dodecyl sulfate. Add diluted sulfuric acid to sodium dodecyl sulfate, heat to reflux, and then titrate with sodium hydroxide standard solution.
Subsequently, prepare the standard solution of sodium dodecyl sulfate whose purity has been determined, and standardize the benzethonium chloride standard solution with titration against sodium dodecyl sulfate standard solution.