Determination of trace chloride ion in copper sulfate solution | Autotitrator COM-A19 - JM Science
HIRANUMA APPLICATION DATA Automatic Titrator Data No. G2 Apr. 5,2019
Metals Determination of trace chloride ion in copper sulfate solution

1. Abstract

This report introduces an example of the determination of trace chloride ion in the solution containing highly concentrated copper sulfate. The precipitation titration with silver nitrate standard solution is generally used for the determination of chloride ion in copper sulfate solution. The endpoint of the titration is detected with indicator method or the potentiometric method. The potentiometric method is used for this sample because the color change of indicator reagent is obscure in this sample containing highly concentrated copper sulfate. However, it tends to show the difficulty to detect the endpoint because of the less sensitivity of electrode under highly concentrated copper sulfate. Therefore the potentiometric titration is performed with the specially treated silver electrode that the silver chloride is coated.

Cl- + AgNO₃ → AgCl + NO₃-  

2. Configuration of instruments and Reagents

(1) Configuration of instruments
  Main unit : Hiranuma Automatic Titrator COM series
    Electrode : Silver electrode AG-311A (Silver chloride-coated electrode)
Reference electrode MS-231Z
*The following electrodes are also usable.
・AGR-801AZ (Silver chloride-coated silver reference combination electrode)
・AGR-811AZ (Silver chloride-coated silver reference combination electrode,double junction type)
*Remark
The general reference electrode (RE-201Z) cannot be used for this titration because KCl inner solution might come out to sample solution and it causes measurement error.
The inner electrodes of AGR-801Z and MS-231Z use mercury (I) sulfate. When these electrodes are disposed, please ask the specialized industrial waste disposal operator.
(2) Reagent    
  Titrant : 0.01 mol/L Silver nitrate standard solution


3. Measurement procedure

(1) Dispense 20 mL of sample into a 100 mL beaker with volumetric pipette.
(2) Add about 20 mL of DI water.
(3) Immerse electrodes and titrate with 0.01 mol/L Silver nitrate standard solution.

4. Measurement conditions and results

Examples of titration conditions

Determination of trace chloride ion in copper sulfate solution Hiranuma Autotitrator COM-A19 from JM Science

Measurement results

Number of
measurement
Size
(mL)
Titrant
volume(mL)
Chloride ion
Concentration (ppm)
1 20 1.294 23.0508
2 20 1.294 23.0508
3 20 1.290 22.9796
Statistic
calculation
Avg. 23.03 ppm
SD 0.04 ppm
RSD 0.18 %

 

Example of titration curve

5. Note

Indicator electrode
  The sensitivity of the electrode is important for this measurement. It tends to show the difficulty on the automatic endpoint detection because of the less sensitivity of normal electrode under highly concentrated copper sulfate. The silver chloride-coated silver electrode can provide acceptable results on this measurement. However, it is degraded by prolonged use even if the silver chloride is coated on the electrode. The recoating of silver chloride is required when the sensitivity of electrode gets lowered.

 

Keywords: Chloride ion, Precipitation titration, Copper sulfate solution, Copper sulfate plating solution,Silver chloride coating

 

*Some measurement would not be possible depending on optional configuration of system.

Hiranuma aquacounter autotitrator com-a19Metals

Featured Instruments

Featured Titrators

View all
Aquacounter AQ-300 Coulometric Karl Fischer Titrator | Hiranuma
Hiranuma Aquacounter AQ-300 Coulometric Karl Fischer Titrator | Hiranuma
Sale price$5,000.00 USD Regular price$6,990.00 USD
Aquacounter AQV-300 Volumetric Karl Fischer Titrator | Hiranuma - JM Science
Hiranuma Aquacounter AQV-300 Volumetric Karl Fischer Titrator | Hiranuma
Sale price$5,000.00 USD Regular price$6,990.00 USD
Aquacounter COM-28S Titrator | HiranumaAquacounter COM-28S Titrator | Hiranuma
Request a quote
Aquacounter COM-28 Titrator | HiranumaAquacounter COM-28 Titrator | Hiranuma
Request a quote