Oil products – Fuel oil | Karl Fischer titrator AQ-300/MOICO-A19/EV-2000L - JM Science
HIRANUMA APPLICATION DATA Karl Fischer Titrator Data No. KF8 Jun.6. 2017
Water contents Oil products – Fuel oil

1. Abstract

 Water content of oil products are determined by Karl Fischer coulometric titrator. In coulometric titration, iodine of Karl Fischer reagent is generated by electrolysis and generated iodine quantitatively reacts with water. Reaction formula is described below.

 H₂O + I₂ + SO₂ + 3RN + CH₃OH → 2RN・HI + RN・HSO₄CH₃

 2RN・HI → I₂ + 2RN + 2H⁺ + 2e⁻

 Generally the fuel oil does not interfere the Karl Fischer reaction and direct injection method could apply. Suitable anode solution is selected for dissolving oil samples.
 It is known that some of the oil additives interfere Karl Fischer reaction. In that case, azeotropic distillation method with Oil evaporator is appropriate. Water is separated from oil sample by distillation and introduced to electrolytic cell with carrier gas.

2. Apparatus and Reagents

(1) Apparatus    
  Titrator : HIRANUMA Karl Fischer Coulometric titrator AQ-series or MOICO-A19
  Evaporator : Oil Evaporator EV-2000L
  Electrolytic cell : Standard Cell
(2) Reagents for direct injection method
  Anode solution : Hydranal coulomat AG-H / Oil
  Cathode solution : Hydranal coulomat CG
(3) Reagents for azeotropic distillation method
  Anode solution : Hydranal coulomat AG (for general use, nonhalogenated)
  Cathode solution : Hydranal coulomat CG
  Distillation solvent : Dehydrated toluene
  Carrier gas : Nitrogen gas

3. Procedure

3.1. Direct injection method

(1) Fill 100 mL of anode solution and one ampoule of cathode solution into the electrolytic cell as shown in Fig.3.1.
(2) Start blanking to attain stable background.
(3) Wash the syringe with sample.
(4) Draw the sample into syringe and then weigh the syringe.
(5) Inject sample from rubber septum of electrolytic cell as shown in Fig.3.2.
(6) Start titration. Measurement parameter is shown in Table 4.1.
(7) Weigh the syringe again and then set the difference of weight to sample size.

Fig.3.1. Preparation of the reagents.



Fig.3.2. Injection of sample.

 

3.2. Azeotropic distillation method

(1) Fill 100 mL of anode solution and one ampoule of cathode solution into the electrolytic cell as shown in Fig.3.1.
(2) Start blanking to attain stable background.
(3) Connect electrolytic cell and evaporation chamber with tube. Flow carrier gas with 50 mL/min
(4) Fill 5 mL of distillation solvent in evaporation chamber and heat the chamber at 120 ℃.
(5) Keep blanking to attain stable background with carrier gas flowing into electrolytic cell.
(6) Wash the syringe with sample.
(7) Draw the sample into syringe and then weigh the syringe.
(8) Inject sample from rubber septum of distillation chamber as shown in Fig.3.3 and Fig.3.4.
(9) Start titration. Measurement parameter is shown in Table 4.2.
(10) Weigh the syringe again and then set the difference of weight to sample size.


Fig.3.3. Schematic diagram of azeotropic distillation method.



Fig.3.4. Sample injection into evaporation chamber.

4. Parameters and results

Table 4.1 Parameters for direct injection method

Condition File    
Cal Mode 0:Sample weight(net)  
  X=(H₂O-BLANK)/SIZE  
Interval Time 20 sec
Current SLOW  
S-Timer 0 min
Blank Value 0 ug
Unit Mode AUTO  
Auto Interval 0 g
Minimum Count 5 ug
Back Ground ON  
Sample Size Input Every Time  
Cell Type Standard  

 

Table 4.2 Parameters for azeotropic distillation method

Condition File    
Cal Mode 0:Sample weight(net)  
  X=(H₂O-BLANK)/SIZE  
Interval Time 40 sec
Current SLOW  
Min.Timer 5 min
Blank Value 0 ug
Unit Mode AUTO  
Auto Interval 0 g
Minimum Count 5 ug
Back Ground ON  
Sample Size Input Every Time  
Cell Type Standard  

 

Table 4.3 Results of water content measurement in Oil products

Sample Apparatus Cell Reagent Sample
Size (g)
Water
(μg)
Water
Content
Gasoline AQ+EV-L Standard AG 2.1268 82.8 39.0 ppm
      CG 2.2044 83.0 37.7
        2.1970 77.6 35.3
             
Heavy oil A AQ+EV-L Standard AG 2.0054 132.3 66.0 ppm
      CG 1.9829 122.3 61.7
        1.9999 130.4 65.2
             
Heavy oil C AQ+EV-L Standard AG 0.9529 15482.0 1.6248 %
      CG 0.2509 3911.3 1.5589
        0.4938 7982.0 1.6164
             
Aviation fuel AQ Standard AG-H 9.5000 398.8 42.0 ppm
      CG 9.5402 388.0 40.7
        9.6370 398.1 41.3
             

5. Note

(1) Use dried syringe and syringe vial for preventive of contamination by atmospheric water.
(2) To measure 100 µg or less water detection, make sure stability of blanking. Low and stable background value is important factor for trace level of water measurement.
(3) Put appropriate anode solution in use according to the solubility of the sample. For example, Hydranal Coulomat AG-H and Oil are suitable for oils with direct injection method.
Note : these reagents does not correspond to Fritless cell.
Hiranuma aquacounter aq-300 coulometric karl fischer titratorHiranuma aquacounter coulometric karl fischer titrator moico-a19Hiranuma aquacounter ev-2000l oil evaporatorWater contents

Featured Instruments

Featured Titrators

View all
Aquacounter AQ-300 Coulometric Karl Fischer Titrator | Hiranuma
Hiranuma Aquacounter AQ-300 Coulometric Karl Fischer Titrator | Hiranuma
Sale price$5,000.00 USD Regular price$6,990.00 USD
Aquacounter AQV-300 Volumetric Karl Fischer Titrator | Hiranuma - JM Science
Hiranuma Aquacounter AQV-300 Volumetric Karl Fischer Titrator | Hiranuma
Sale price$5,000.00 USD Regular price$6,990.00 USD
Aquacounter COM-28S Titrator | HiranumaAquacounter COM-28S Titrator | Hiranuma
Request a quote
Aquacounter COM-28 Titrator | HiranumaAquacounter COM-28 Titrator | Hiranuma
Request a quote