Drugs and Medicines – KF Volumetry, Back-Titration, Dextromethorphan Hydrobromide Monohydrate | Karl Fischer titrator AQV-300/MOIVO-A19 - JM Science
HIRANUMA APPLICATION DATA Karl Fischer Titrator Data No. KF11 Apr. 13, 2022
Water contents Drugs and Medicines – KF Volumetry Back-Titration Dextromethorphan Hydrobromide Monohydrate

1. Abstract

 Water contents of drugs and medicines could be determined by Karl Fischer volumetric titrator. In volumetric titration, the titrant have a factor which means the ability to react with how many milligrams of water per 1 mL of titrant. Factor is pre-determined before sample measurement and water content of sample is calculated from consumed titrant volume within sample measurement.
 There are two methods in KF volumetric titration, forward-titration and back-titration. Forward-titration is used in general. However, some of the sample should be measured by back-titration method according to Pharmacopeias. In back-titration method, firstly excess amount of KF titrant is added to the water contained in the sample, and then the remaining unreacted of KF titrant is titrated with water-in-methanol solution. Therefore, it is necessary to measure the factor not only KF titrant but also the water-in-methanol before the sample measurement.
 An example for water contents measurements of dextromethorphan hydrobromide monohydrate performed by back-titration method is introduced here. The measurement method was determined with the reference to Japanese Pharmacopeia. Please refer to Application Data No. 22 for the suitability test of Japanese Pharmacopoeia eighteenth edition by the back titration method.
Reference
1) Japanese Pharmacopoeia Eighteenth Edition

2. Apparatus and Reagents

(1) Apparatus    
  Titrator : HIRANUMA Karl Fischer Volumetric titrator AQV-series or MOIVO-A19
  Additional Buret : 1 unit
  Titration cell : Standard cell without drain valve
  Sampler : Syringe
Powder funnel (Outer diameter of leg less than 12 mm)
(2) Reagents    
  KF Titrant : AQUALYTE KF3 (HIRANUMA)
An alternative if it is difficult to prepare, HYDRANAL Composit 5 (Honeywell)
  Water-in-methanol
  solution
: 2 mg/mL, prepared from reagent grade methanol and DI water
  Titration solvent : HYDRANAL Methanol Dry (Honeywell)
(3) Sample    
    Dextromethorphan Hydrobromide Monohydrate (Guaranteed reagent)

3. Procedure

 3.1. Factor determination of KF titrant

(1) Fill 50 mL of titration solvent into the titration cell as shown in Fig.3.1.
(2) Start blanking to attain stable background.
(3) Draw DI water into syringe. Then, weigh the syringe and record its read (Size 1 [g]).
(4) Inject a few drops of water from rubber septum of titration cell as shown in Fig.3.2. Approximate amount of addition is 20-40 mg. (40-60 mg when using Composite 5 for KF titrant)
(5) Start titration. Measurement parameter is shown in Table 4.1.
(6) Weigh the syringe again and record its read (Size 2 [g]). The difference of (Size 1-Size 2 [g]) is set as sample size.
(7) Repeat the measurement 3 times and obtain an average value for factor of KF titrant.

Fig.3.1 Preparation of the reagents



Fig.3.2 Injection of sample

 

3.2. Factor determination of water-in-methanol solution

(1) Fill 50 mL of titration solvent into the titration cell as shown in Fig.3.1.
(2) Start blanking to attain stable background.
(3) Start titration. Measurement parameter is shown in Table 4.2.
10 mL of KF titrant is dispensed and then titrated by water-in-methanol solution.
(4) Repeat the measurement 3 times and obtain an average value for factor of water-in-methanol solution.

 

3.3. Sample measurement by back-titration

(1) Fill 50 mL of titration solvent into the titration cell as shown in Fig.3.1.
(2) Start blanking to attain stable background.
(3) Put a sample container, powder funnel and spoon on the balance. Record its read (Size 1 [g]).
(4) Open the glass stopper of titration cell lid to introduce the sample with powder funnel as shown in Fig.3.3.
(5) Start titration. First, 10 mL of KF titrant is dispensed into the cell and then titrated with water-in-methanol solution. Measurement parameter is shown in Table 4.3. Set the factors measured in section 3.1 and 3.2 for the KF reagent and the water-in-methanol solution.
(6) Weigh the sample container, powder funnel and spoon again and record its read (Size 2 [g]). The difference of (Size 1-Size 2 [g]) is set as sample size.

Fig.3.3 Introduction of sample with powder funnel

 

3.4. Blank test measurement by forward-titration

(1) Fill 50 mL of titration solvent into the titration cell as shown in Fig.3.1.
(2) Start blanking to attain stable background.
(3) Open the glass stopper of titration cell lid and introduce the funnel. Without adding the sample, leave to stand for the same amount of time as it takes to add the sample in section 3.3, then attach the stopper.
(4) Start titration. Blank test is performed with forward-titration. Measurement parameter is shown in Table 4.4.
(5) Set the result of titrant volume obtained in the blank test as the blank value for the measurement conditions of the sample.

4. Parameters and results

Table 4.1 Parameters for factor measurement of KF titrant

Item    
Cal Mode 7:KF Factor  
  (By Pure water)  
Interval Time 30 sec
Max Volume 20 mL
Min Feed Vol. 0.01 mL
S.Timer 0 min
KF Buret No. 1  
KF Speed(OUT) 24 mL/min
KF Speed(IN) 24 mL/min
Back Ground OFF  
Item    
     
     
Sample Size Input Every Time  
Blank Value 0 mL
E.P Detection uA  
Solvent S,O,CE  
C.P Level 150 uA
E.P Level 200 uA
Auto Interval 0 g
Auto Input OFF  

 

Table 4.2 Parameters for factor measurement of water-in-methanol solution

Item    
Cal Mode 10:MeOH Factor  
  (Back tit.MeOH)  
Interval Time 30 sec
Max Volume 20 mL
Min Feed Vol. 0.01 mL
S.Timer 0 min
KF Factor 2.9263 mg/mL
KF Buret No. 1  
KF Disp.Vol. 10 mL
MeOH Buret No. 2  
MeOH Disp.Vol. 10 mL
MeOH Timer 1 min
     
Item    
     
     
KF Speed(OUT) 24 mL/min
KF Speed(IN) 24 mL/min
MeOH Speed(OUT) 24 mL/min
MeOH Speed(IN) 24 mL/min
Back Ground OFF  
Blank Value 0 mL
E.P Detection uA  
C.P Level 150 uA
E.P Level 200 uA
Fact Meas. 1  
Auto Input OFF  

 

Table 4.3 Parameters for sample measurement by back-titration method

Item    
Cal Mode 11:Back tit1  
  (Fixed KFreagent disp)  
Interval Time 30 sec
Max Volume 20 mL
Min Feed Vol. 0.01 mL
S.Timer 1 min
KF Factor 2.9263 mg/mL
KF Buret No. 1  
KF Disp.Vol. 10 mL
KF Timer 1 min
MeOH Factor 2.0279 mg/mL
MeOH Buret No. 2  
MeOH Disp.Vol. 0 mL
MeOH Timer 0 min
Item    
     
     
KF Speed(OUT) 24 mL/min
KF Speed(IN) 24 mL/min
MeOH Speed(OUT) 24 mL/min
MeOH Speed(IN) 24 mL/min
Back Ground OFF  
Sample Size Input Every Time  
Blank Value 0.01 mL
Unit Mode AUTO  
E.P Detection uA  
C.P Level 150 uA
E.P Level 200 uA
Auto Interval 0 g

 

Table 4.4 Parameters for blank test by forward-titration method

Item    
Cal Mode 13:Auto input blank value mode  
     
Interval Time 30 sec
Max Volume 20 mL
Min Feed Vol. 0.01 mL
S.Timer 0 min
     
     
     
     
Item    
     
     
KF Buret No. 1  
KF Speed(OUT) 24 mL/min
KF Speed(IN) 24 mL/min
Back Ground OFF  
E.P Detection uA  
C.P Level 150 uA
E.P Level 200 uA
Auto Input OFF  

 

Table 4.5 Results of factor measurement of KF titrant

Sample Meas.
No.
Meas.
Time
Sample
size (g)
Titrant
volume (mL)
KF factor
(mg/mL)
Statistics result
Water 1 0:04:25 0.0490 16.74 2.9271 Avg. 2.9263 mg/mL
  2 0:04:09 0.0520 17.78 2.9246 SD 0.0014 mg/mL
  3 0:03:28 0.0454 15.51 2.9271 RSD 0.05 %

 

Table 4.6 Results of factor measurement of water-in-methanol solution

Sample Meas.
No.
Meas.
Time
Sample
size (mL)
Titrant
volume (mL)
KF factor
(mg/mL)
Statistics result
KF titrant 1 0:05:25 0 14.44 2.0265 Avg. 2.0279 mg/mL
10 mL 2 0:08:03 0 14.44 2.0265 SD 0.0024 mg/mL
  3 0:10:18 0 14.41 2.0307 RSD 0.12 %

 

Table 4.7 Results of sample measurement

Sample Meas.
No.
Meas.
Time
Sample
size (g)
Titrant
volume (mL)
Water
(mg)
Water
content (%)
Statistics result
Blank test 1 0:00:40 0.01 Avg. 0.01 mL
  2 0:00:36 0.01    
Dextromethorphan 1 0:07:02 0.2000 9.62 9.725 4.8625 Avg. 4.9106 %
hydrobromide 2 0:09:12 0.1968 9.64 9.685 4.9212 SD 0.0438 %
monohydrate 3 0:12:57 0.1990 9.56 9.847 4.9482 RSD 0.89 %

5. Note

(1) Sampling tools should be dried up well before use.
(2) Purge and fill the titrant homogeneously into the buret.
(3) Organic solvents which have relatively high volume expansion coefficient are used as a constituent of titrant. For accurate measurement, factor titration and sample measurement should be performed at the same room temperature as much as possible.
(4) When adding a solid sample to the cell by opening the stopper of the cell lid, blank test should be performed separately to measure the water incorporated during the procedure of sample addition. Perform the blank test by forward-titration method instead of back-titration, and set the obtained titration volume in the blank value of the measurement condition of the sample.
(5) Compared to forward-titration, back-titration tends to take longer time to measure. If the titration volume of the water-in-methanol solution can be predicted in advance, set a value 3 to 5 mL smaller than the expected titration volume in the parameter “MeOH Disp. Vol.”, and also set the parameter “MeOH Timer” to 1 minute. As a result, the measurement time is shortened by dispensing the water-in-methanol solution with the set value immediately after the start of titration with the water-in-methanol solution.

Keywords: Karl Fischer, Volumetric titration, Back titration, Pharmacopeia, Dextromethorphan hydrobromide monohydrate
Hiranuma aquacounter aqv-300 volumetric karl fischer titratorHiranuma aquacounter volumetric karl fischer titrator moivo-a19Water contents

Featured Instruments

Featured Titrators

View all
Aquacounter AQ-300 Coulometric Karl Fischer Titrator | Hiranuma
Hiranuma Aquacounter AQ-300 Coulometric Karl Fischer Titrator | Hiranuma
Sale price$5,000.00 USD Regular price$6,990.00 USD
Aquacounter AQV-300 Volumetric Karl Fischer Titrator | Hiranuma - JM Science
Hiranuma Aquacounter AQV-300 Volumetric Karl Fischer Titrator | Hiranuma
Sale price$5,000.00 USD Regular price$6,990.00 USD
Aquacounter COM-28S Titrator | HiranumaAquacounter COM-28S Titrator | Hiranuma
Request a quote
Aquacounter COM-28 Titrator | HiranumaAquacounter COM-28 Titrator | Hiranuma
Request a quote