Description
The Aquacounter AQS-22320AS automated solid evaporator system is an automated Karl Fischer titrator system consisting of the AQ-2200 (coulometric titrator) and the EV-2000 (solid evaporator) units.
The system is designed to efficiently measure the moisture in solid and powdered samples such as various resin pellets, medicines (powdered and solid), various inorganic powders and granules, copy machine toner, etc. The automated system allows the analyst to load the samples into the glass vials and then place the glass vials on the sample tray of the system. The operator only needs to push the “start” button once to begin the analyses.
Features
- Sample changer holds up to 20 vials.
- Equipped with Automatic KF reagent exchanging function.
- Wide temperature setting covers the full range of azeotropic points.
- Reduces costs for KF reagent and waste disposal.
- Can process large numbers of samples.
- Manufactured by Hiranuma in Japan
ITEM |
Solid Evaporator SE-320 |
Number of sample vials |
20 positions |
Volume of sample vial |
Max. 10mL |
Seal |
PTFE cap with Aluminum foil |
Carrier gas |
Nitrogen gas* *Optional N2 gas generator from JMS is recommended |
Carrier gas flow rate |
30 - 300mL/min |
Carrier gas desiccant |
Molecular Sieves |
Heater |
Aluminum block |
Heating temperature |
Room temperature to 300ÅãC |
Power supply / Power consumption |
AC100 - 240V, 50/60Hz, 150VA |
Dimensions |
Sample changer + heater unit: 350 (W) x 510 (D) x 390 (H) mm Valve unit: 275 (W) x 440 (D) x 280 (H) mm |
Weight |
Sample changer + heater unit: Approx. 19kg Valve unit: Approx. 9kg |