Chromaster 5410 UV Detector | Hitachi Brand


Request a quote

Description

The Chromaster 5410 UV Detector from Hitachi offers powerful chromatography performance and convenience, thanks to its low dwell volume and carry-over, coupled with highly sensitive detector technology. Its modular design and versatile accessories make it suitable for a variety of applications, from robust industrial processes to ultra-sensitive pharmaceutical analysis.

Features

Low noise, low drift, and a high sensitivity detection

A noise level 0.5 × 10 AU*² (or less) can be achieved, allowing better sensitivity than ever. With a low drift of 1.0 × 10⁻⁴ AU / hr*³ (or less), these detectors deliver excellent baseline stability.
*2, 3 Under a specified condition

Two-wavelength simultaneous measurement function

The two-wavelength detection function* permits measurements at short data acquisition intervals of 400 ms* and 800 ms per wavelength. The result is chromatograms with sharp peak shapes.
*4 Controlled by CDS only
*5 400 ms is available only if the wavelength interval is 160 nm or less.

Hitachi UV Detector 5410 two-wavelength simultaneous analysis data

Included Accessories

  • AC Adapter
  • Data Cable

Product Specifications

Item Specifications
Optical system Double-beam ratio photometric system
Light source D2 lamp, Hg lamp for checking wavelength
Wavelength range 190 nm to 600 nm
Wavelength accuracy ±1 nm
Spectral bandwidth 6 nm
Noise
≤−0.5 × 10 − 5 AU at 250 nm, under a specified condition
Drift
≤−1.0 × 10 − 4 AU/h at 250 nm, under a specified condition
2-wavelength measurement 2 wavelengths in wavelength regions 190 to 350 nm and 351 to 600 nm, respectively (Minimum wavelength interval 5 nm, max. wavelength interval 160 nm with data sampling period set at 400 ms)
Response 0.01, 0.02, 0.05, 0.1, 0.5, 1, 2 sec
Materials of wetted parts Quartz glass, Fluororesin, Stainless steel
Functions of GLP (a)D2 lamp/Hg lamp lighting time, lighting number of times, and replacement record (b)Key lock (c)D2 lamp energy check and D2 lamp wavelength check (d)Hg lamp wavelength check
Flow cell 13 μL (Optical path length 10 mm)
Thermostatically flow cell
Optional, Environmental temperature range: 4 to 30°C
Dimensions and weight 340 (W) × 440 (D) × 140 (H) mm, Approx.13 kg
Power supply and Power consumption DC24 V, 2.5 A (Maximum)/60 W (power supply from organizer) *Please purchase the AC adaptor (150 W) when there is no organizer.

 